Clinical Performance Summary - Defibrillation

M3500B HeartStart XLT Defibrillator/Monitor

13-15

13

Conclusion

The 150J SMART Biphasic waveform defibrillated at higher rates than 200-
360J monophasic waveforms, resulting in more patients achieving return of
spontaneous circulation (ROSC) (p=0.01). EMS system outcomes of survival
discharge were not significantly different statistically. However, patients
resuscitated with the lower energy SMART Biphasic waveform were more
likely to have good cerebral performance (CPC, cerebral performance cate-
gory) (p=0.04).

Table 13-3 Clinical Summary - Defibrillation

Biphasic Patients

Number/(Percent)

Monophasic Patients

Number/(Percent)

P Value

(chi-square)

Defibrillation Efficacy

Single shock only
<2 shocks
<3 shocks

52/54 (96%)
52/54 (96%)
53/54 (98%)

36/61 (59%)
39/61 (64%)
42/61 (69%)

<0.0001
<0.0001
<0.0001

Patients Defibrillated

54/54 (100%)

49/58 (84%)

0.003

Return of Spontaneous
Circulation

41/54 (76%)

33/61 (54%)

0.01

Survival to Hospital
Admission

33/54 (61%)

31/61 (51%)

0.27

Survival to Hospital Dis-
charge

15/54 (28%)

19/61 (31%)

0.69

CPC = 1 (Good)

13/15 (87%)

10/19 (53%)

0.04