Pacing, Pacing -4, Nd “pacing” ch – Philips HEARTSTART XLT M3500B User Manual

Page 86

Advertising
background image

Pacing

8-4

Pacing

Pacing

To perform pacing:

1. Press

. The green LED next to

lights up and the dia-

logue box appears.

Figure 8-1 Pacing Display

The message

6WRS indicates that the pacing function is on but paced pulses

are not being delivered. The pacer turns on in the mode last used.

2. Verify that the dot markers appears near the middle of the QRS complexes

of the ECG.

If the dot markers do not appear, or are in the wrong location, adjust the
ECG size or select another lead. (See “Monitoring the ECG” in Chapter
4.)

The ECG will only appear if the 3- or 5-lead monitoring cable and elec-
trodes are used.

3. Press

to change to fixed mode, if R-wave detection is unreliable.

The pacing dialogue box displays the current mode. To switch back to

demand mode, press

again.

Pacer

Pacer

ENERGY

SELECT

HR ALARM

LEAD

SELECT

CHARGE

Shocks:0

00:00:8

78

Lead II

SpO2

150 J

Pacer Stop

Demand Mode

70ppm 30mA

Mode

Mode