Marking Events

9-2

Storing, Retrieving & Printing

Marking Events

The

button allows you to annotate the ECG strip at the point in time the

button is pressed. In AED Mode, when monitoring is disabled, the event is

marked with a

. In Manual Mode, or when monitoring is enabled in AED

Mode, you use the softkeys to select the annotation from the choices dis-
played (See Figure 9-1)*. If no selection is made, the event is marked with a

.

Figure 9-1 Annotations

NOTE

*In Australia and the U.K.,

(3, is replaced by $'51 (adrenaline).

The marked event is stored in the Event Summary. If the printer is configured

to

3ULQW RQ 0DUN, an ECG strip prints when

is pressed. If the printer is

configured to

 VHFRQG GHOD\, the strip is 9 seconds and includes 6 seconds pre-

ceding the event and 3 seconds following the event. If

1R 'HOD\ is configured, a

3 second ECG strip prints in real time. To stop printing before the entire strip
is printed, press

.

Mark

Event

EPI

ATRO

LIDO

OTHER

Pads

Shocks:0

00:00:8

72

CHARGE

ENERGY

SELECT

150 J

Mark

Event

Print

Strip