Pacing

M3500B HeartStart XLT Defibrillator/Monitor

8-5

8

4. Adjust the rate to the desired number of paced pulses per minute (ppm).

Press up,

, or down,

, on

to increase or decrease the num-

ber of paced pulses per minute.

5. Press

to start pacing.

The message

3DFLQJ indicates that paced pulses are being delivered in the

selected mode at the rate and output level displayed.

Figure 8-2: Pacing Display

NOTE

Pacing will not start if there is a problem with the multifunction defib elec-
trode pads connections or, in demand mode, a problem with the ECG monitor-
ing electrodes connections. Should a problem exist, a system message is
displayed. Be attentive to system messages.

Rate

Start

Stop

ENERGY

SELECT

HR ALARM

LEAD

SELECT

CHARGE

Shocks:0

00:00:8

78

SpO2

150 J

Pacing

Demand Mode

70ppm 30mA

Lead II