Pausing for CPR

3-12

Defibrillating in AED Mode

Pausing for CPR

At the completion of a shock series or when no shock is advised, the Heart-
Start XLT prompts you to

&KHFN 3DWLHQW &KHFN 3XOVH . It allows eight seconds for

you to check the pulse, then prompts you as follows:

Figure 3-10 Press Pause Display

If CPR is needed, press

. While paused, the Pause Timer indicates the

elapsed time and the total duration of the Pause state, in seconds. The Pause
Timer is configurable to meet your local CPR protocol needs. Rhythm, SpO

2

and heart rate monitoring alarms are suspended for the duration of the pause.

NOTE

* This section describes how the Pause state functions using the default con-
figuration. If your HeartStart XLT is configured to support the European
Resuscitation Council Guidelines for Resuscitation, refer to the “ERC Proto-
col” se
ction on page 3-16 for details.

Shocks: 1

00:00:50

If Needed

Press PAUSE

And Begin CPR

PAUSE

ANALYZE