Philips HEARTSTART XLT M3500B User Manual

Page 60

Advertising
background image

Troubleshooting

4-8

Monitoring the ECG

Poor ECG signal
quality

z

The monitoring elec-
trodes are not making
proper contact with the
patient.

z

The monitoring elec-
trodes are outdated or
dried-out.

z

Radio frequency inter-
ference (RFI) is causing
artifact.

z

Check that the monitoring
electrodes are properly
applied. If necessary, pre-
pare the patient’s skin and
apply new electrodes.

z

Check the date code on
the electrodes. Do not
open the electrode pack-
age until immediately
prior to use.

z

Relocate or turn off equip-
ment that may be causing
RFI.

QRS beeper inau-
dible or beeps do
not occur with
each QRS com-
plex.

z

The QRS beeper is con-
figured to Off.

z

The amplitude of the
QRS complex is too
small to detect.

z

Check that the QRS
beeper is configured to
On.

z

Adjust the volume.

z

Adjust the size of the
ECG.

Table 4-5 Troubleshooting Tips (Continued)

Situation

Cause

Solution