System Messages

M3500B HeartStart XLT Defibrillator/Monitor

12-5

12

Momentary Messages

Momentary messages are temporary and only appear on the display for a few
seconds. Each message is accompanied by a three second beep to alert you.
Table 12-2 lists momentary messages.

Table 12-2 Momentary Messages

Message

Possible Cause

Corrective Action

$WWDFK 3DGV

The multifunction defib electrode pads
are not making proper contact with the
patient.

z

Check the pads are applied to the
patient, as directed on the pads’ pack-
age.

z

Replace the pads if the prompt contin-
ues.

$WWDFK /HDGV

An attempt was made to begin pacing in
Demand Mode without ECG leads
attached to the patient.

Attach leads to patient.

'HILE 'LVDUPHG

z

The pads connection is compromised.

z

The mode is changed from Manual to
AED while the defibrillator is
charged.

z

is not pressed within 30 sec-

onds of the defibrillator being
charged.

z

is pressed.

z

Check the pads are applied to the
patient properly.

z

If a shock is indicated, deliver the
shock before changing modes.

z

To deliver a shock, press

within 30 seconds of the defibrillator
being charged.

1R 6KRFN 'HOLYHUHG

Patient impedance is too high.

z

Make sure the pads are applied prop-
erly.

z

Replace the pads, if necessary

&KHFN 3ULQWHU

Printer paper is absent or jammed; the
printer door is not closed properly.

z

Reload printer paper.

z

Make sure the door is closed properly.

6+2&.

',6$50

6+2&.