Monitoring, Monitoring -7 – Philips HEARTSTART XLT M3500B User Manual

Page 67

Advertising
background image

Monitoring

M3500B HeartStart XLT Defibrillator/Monitor

5-7

5

Monitoring

To monitor SpO

2

:

1. If the HeartStart XLT is not on, press On.

2. Apply the appropriate sensor to the patient.

3. Make sure the sensor cable is connected to the HeartStart XLT.

4. Press

to turn on SpO

2

monitoring.

A dashed line (---) is displayed under

, while the oxygen saturation is

measured and an SpO

2

value is calculated. In a few seconds the SpO

2

value is

displayed in place of the dashed line. As the patient’s oxygen saturation
changes, the SpO

2

value is updated continuously.

Figure 5-3 SpO

2

Monitoring Display

To the right of the SpO

2

value, a pleth bar and SpO

2

alarm indicator are dis-

played. The pleth bar should be observed for fluctuation. It is an indication of
pulsation detected by the sensor. The pleth bar should not be used as the sole
indicator of pulsation because it can be influenced by movement and artifact.

The

symbol indicates no alarm is set.

Below the SpO

2

value is the pulse rate derived from the pulse oximetry.

6S2



6S2



PAUSE

HR ALARM

LEAD

SELECT

ANALYZE

Shocks: 3

00:00:49

Monitoring Rhythm

78

Pads

SpO2

98

PR 62