2250 vertical total too small, 2270 vertical total is even, 2280 vertical total too large – Quantum Data 801GC User Manual

Page 319: 2300 vertical active too small, 2310 vertical blanking too small, 2320 vertical active too small

Advertising
background image

8-11

2250

Vertical total too small

There must be at least a total of 5 horizontal scan lines per
frame with interlaced operation.

■

Increase the vertical period so that it’s greater than 4
lines.

■

Switch to non-interlaced operation.

2270

Vertical total is even

There must be an odd numbered total of horizontal scan
lines per frame with interlaced operation.

■

Change the vertical period (in lines) to an even number.

■

Switch to non-interlaced operation.

2280

Vertical total too large

The generator does not support the given vertical period.

■

Decrease the vertical period to less than 4096 lines
per frame if non-interlaced.

■

Decrease the vertical period to less than 4097 lines
per frame if interlaced.

2300

Vertical active too small

There must be at least 1 active horizontal scan line with
non-interlaced operation.

■

Increase vertical active so that it’s greater than 1 line.

2310

Vertical blanking too small

There must be at least 1 line of vertical blanking with non-
interlaced operation.

■

Increase the vertical period.

■

Decrease vertical active.

2320

Vertical active too small

There must be at least 2 active horizontal scan lines with
interlaced operation.

■

Increase vertical active so that it’s greater than 2 lines.

■

Switch to non-interlaced operation.