2490 eq before too large, 2495 eq after too large, 2496 eq after too large – Quantum Data 801GC User Manual

Page 323: 2704 horizontal physical size too small, 2705 invalid physical size units

Advertising
background image

8-15

2490

EQ before too large

The interval during which pre-equalization pulses occur
cannot be greater than the vertical sync pulse delay.

Decrease EQ Before.

Increase the vertical sync pulse delay.

2495

EQ after too large

The interval during which post-equalization pulses occur
cannot be greater than the interval between the end of the
vertical sync pulse and the end of the shorter vertical blanking
interval between fields with interlaced operation.

Decrease EQ After.

Decrease the vertical sync pulse width.

Decrease vertical sync pulse delay.

Increase the vertical period (in lines).

Decrease the vertical active.

2496

EQ after too large

The interval during which post-equalization pulses occur
cannot be greater than the interval between the end of the
vertical sync pulse and the end of the vertical blanking.

Decrease EQ After.

Decrease the vertical sync pulse width.

Decrease vertical sync pulse delay.

Increase the vertical period (in lines).

Decrease the vertical active.

2550

Not enough video memory

The generator does not have enough memory space to store
the total number of active pixels required for one picture.

Decrease the vertical active.

Decrease the horizontal active.

2704

Horizontal physical size too small

The physical horizontal size must be greater than zero.

Change the horizontal physical size to a positive,
nonzero, value.

2705

Invalid physical size units

An invalid physical size unit of measure exists in the file.

Reedit both physical sizes.