2330 vertical blanking too small, 2350 vertical pulse too small, 2370 vertical pulse too large – Quantum Data 801GC User Manual

Page 320: 2390 vertical pulse too large

Advertising
background image

8-12

Chapter 8: Error Messages

Model 801GC, 801GF & 801GX¥Rev. A

2321

Vertical active must be even when in inter-
laced scan mode

The generator does not support vertical active periods (in
lines) that cannot be evenly divided by two (2) when
interlaced operation is selected.

■

Make the vertical active period (in lines) evenly divisible
by 2.

■

Change to progressive (non-interlaced) scan mode.

2330

Vertical blanking too small

There must be at least 3 lines of vertical blanking with
interlaced operation.

■

Increase the vertical period.

■

Decrease vertical active.

■

Switch to non-interlaced operation.

2350

Vertical pulse too small

The generator does not support a vertical sync pulse width
value of less than one scan line in duration. The firmware
makes the pulse a half-line shorter than the entered value
when European type sync is selected.

■

Increase the vertical sync pulse width.

2370

Vertical pulse too large

The generator does not support a vertical sync pulse width
greater than the number of blanked scan lines per frame
with non-interlaced operation.

■

Decrease the value of vertical sync pulse width.

■

Increase the value of vertical period (in lines).

■

Decrease vertical active.

2390

Vertical pulse too large

The generator does not support a vertical sync pulse width
greater than the least number of blanked scan lines ((Vper
- Vact - 1)

÷

2 ) between fields with interlaced operation.

■

Decrease the value of vertical sync pulse width.

■

Increase the value of vertical period (in lines).

■

Decrease vertical active.