Tektronix AWG610 User Manual

Page 111

Advertising
background image

Tutorials

AWG610 Arbitrary Waveform Generator User Manual

2-73

You will create the sequence list shown in Table 2–18. This sequence is used as a
subsequence and is called from the main sequence that you create in Create the
Main Sequence
on page 2–75. This sequence runs as follows:

1. Line 1: outputs the gaussian noise waveform 40,000 times and then goes to

line 2.

2. Line 2: outputs the ramp waveform 60,000 times and then goes to the next

line (3).

3. Line 3: outputs the triangle waveform 60,000 times and then goes to the

next line (4).

4. Line 4: output the sine waveform 30,000 times and then quits the subse-

quence and returns to the main sequence.

Table 2-18: Sequence table contents in SUBSEQ.SEQ

Line CH1

CH2

Repeat

Count

Wait

Trigger

Goto

One

Logic

Jump

1

GAUSSN.WFM

40,000

2

RAMP.WFM

60,000

3

TRIANGLE.WFM

60,000

4

SINE.WFM

30,000

In the sequence file used as subsequence, the Wait Trigger, Goto One and Logic
Jump are neglected. They are effective only in the main sequence.

Do the following steps to create the subsequence:

1. Push the Data Entry bottom button.

2. Push the Insert Line side button.

This displays the line number in the Line column and allows you to edit the
line.

3. Push the Enter file name side button.

The dialog box listing files appears at the center of the screen.

4. Select gaussn.wfm from the dialog box using the general purpose knob.

5. Push the OK side button.

The waveform file name gaussn.wfm appears in the CH1 column.

6. Push the

b button once to move the highlighted cursor to the next line.

Create the Subsequence

Advertising