Appendix B: Performance Verification

B-40

AWG610 Arbitrary Waveform Generator User Manual

Trigger Input Tests

These procedures check the trigger level accuracy of the AWG610 Arbitrary
Waveform Generator.

Equipment

required

Two 50ĂW BNCĂcoaxial cables, aĂ50ĂW SMAĂcoaxial cable,

ĂaĂSMA(Fe)ĆBNC(Ma)Ăadapter,ĂaĂBNCĆTĂ(male to 2 females) adapter , a

function generator, and an oscilloscopeĂ(TDS700).

Prerequisites

The

AWG610 Arbitrary Waveform Generator

must meet the

prerequisites listed on page B-8.

Do the following steps to install the test hookup and set the test equipment
controls:

1. Use a 50

W SMA coaxial cable and an SMA(Fe)-BNC(Fe) adapter to

connect the AWG610 Arbitrary Waveform Generator CH1 output connector
to the oscilloscope CH1 input connector.

2. Use a BNC coaxial cable to connect the function generator to the BNC-T

adapter which goes to the Trigger Input on the AWG610. Connect a 50
W BNC coaxial cable to the BNC-T adapter .Connect the opposite end of the
BNC cable to the CH2 input on the oscilloscope.

Function Generator (AFG310)

BNC T

Adapter

AWG610 Arbitrary Waveform

Generator rear panel

Output

CH2 input

CH1 input

CH1 output

OscilloscopeĂ(TDS700)

SMA(Female)ĆBNC

(Male) adapter

50 W SMAĂcoaxial cable

50 W BNCĂcoaxial cable

50 W BNCĂcoaxial cable

Figure B-15: Trigger input initial test hookup