Change the generator mode

The instrument initializes in the AWG mode when powered on.

Do the following to change the generator mode from AWG to FG :

1. Push SETUP (front-panel)

!Waveform/Sequence (bottom)!

Ez FG... (side) button.

The instrument displays the FG mode screen.

Do the following to change the generator mode from FG to AWG :

1. Push AWG... (bottom) button.

The instrument returns to the AWG mode.

To FG

mode

To AWG

mode

AWG SETUP screen

FG mode screen

Figure 3-73: Change the generator mode

NOTE. All the parameters on the FG mode menu are independent of the AWG
mode parameters. Therefore, the output parameters, such as frequency,
amplitude and offset, have no effect on the parameters set with the SETUP menu
while in the other mode.
In FG mode, the AWG 610 runs CONTINUOUS mode only.

AWG mode to FG mode

FG mode to AWG mode