Appendix B: Performance Verification

AWG610 Arbitrary Waveform Generator User Manual

B-29

Amplitude, Offset Accuracy and Rise Time Tests (Direct DA Out)

These procedures check the accuracy of the AWG610 Arbitrary Waveform
Generator direct waveform outputs; amplitude and offset.

Equipment

required

A 50ĂW BNCĂcoaxial cable, a 50ĂWĂprecision terminator, an

SMA(Fe)-BNC(Ma) adapter,

BNC (female)ĆtoĆdual banana

adapter, and a digital multimeter (DMM).

Prerequisites

The AGW610 Arbitrary Waveform Generator must meet the

prerequisites listedon page B-8.

Do the following steps to install the test hookup and set the test equipment
controls:

1. Use a 50

W

BNC coaxial cable, an SMA(Fe)-BNC(Ma) adapter, a 50

W

precision terminator, and a BNC-to-dual banana adapter to connect the
AWG610 Arbitrary Waveform Generator CH1 output to the DMM input
connector (see Figure B–10).

ББ

ББ

DMM

50 W BNCĂcoaxial cable

InputĂconnector

+ĂBNCĆtoĆdual banana adapter

+Ă50 W BNCĂcoaxialĂcable

AWG610 Arbitrary Waveform Generator

CH1ĂoutputĂconnector

+ĂSMA(Fe)ĆBNC(Ma) adapter

+ĂBNC 50 WĂprecision terminator

+Ă50 WĂBNC coaxial cable

Figure B-10: Direct DA output amplitude accuracy initial test hookup

2. Set the DMM controls as follows:

Mod e . . . . . . . . . . . . . . . . . . . . . . . . . . VDC
Range . . . . . . . . . . . . . . . . . . . . . . . . . 2 V
Input . . . . . . . . . . . . . . . . . . . . . . . . . . Front

Check Amplitude

and

DC Offset