Marker delaytests – Tektronix AWG610 User Manual

Page 421

Advertising
background image

Appendix B: Performance Verification

AWG610 Arbitrary Waveform Generator User Manual

B-61

Marker DelayTests

These procedures check the marker delay function of the AWG610 Arbitrary
Waveform Generator.

Equipment

required

Two 50ĂW SMA coaxial cables, twoĂSMA(Fe)ĆBNC(Ma) adapters, two

Ă50ĂW SMA terminatorsĂ andan oscilloscope.

Prerequisites

The AWG610 Arbitrary Waveform Generator must meet the

prerequisites listedon page B-8.

NOTE. Connect two 50 W SMA terminators to the each inverted marker out-
put connectors, during the marker output tests.

NOTE. Two 50 W SMA coaxial cables must have same length.

Do the following steps to install the test hookup and set the test equipment
controls:

1. Use the 50

W

SMA coaxial cables and SMA-to-BNC adapters to connect the

AWG610 Arbitrary Waveform Generator MARKER1 and MARKER2
outputs to the oscilloscope CH1 and CH2 input connectors (see
Figure B–32).

2. Connect a 50

WĂSMA terminator to the MARKER1 and a 50 WĂSMA termi-

nator to the MARKER2 connectors.

ББ

ББ

50 W SMA coaxial cable

50 W SMA coaxial cable

AWG610 Arbitrary Waveform Generator

ConnectĂ50 WĂSMA

terminators to the MARKER1

andMARKER2 connector

OscilloscopeĂ(TDS700)

SMA(Fe)ĆBNC(Ma)

adapters

Figure B-32: Marker delaytest hookup

Advertising