Tektronix AWG610 User Manual

Page 391

Advertising
background image

Appendix B: Performance Verification

AWG610 Arbitrary Waveform Generator User Manual

B-31

11. Follow the substeps below to check the DC offset:

a. Push the FORCE EVENT button.
b. Check that the reading from the oscilloscope display is about

0 V

±

10 mV.

12. Do the following substeps to change the connection to check the CH1:

a. Push the CH1 output button.
b. Disconnect the adapter, terminator and cable from the CH1 output con-

nector.

c.

Connect the adapter, terminator, and cable (removed in step
b) to the CH1 output connector.

d. Push the CH1 output button to turn on the CH1 LED.

13. Repeat steps 8 through step 11 for the AWG610 Arbitrary Waveform

Generator CH1.

14. Push the CH1 output button to turn off the CH1 LED.

15. Push the RUN button to turn off the RUN LED.

16. Disconnect the DMM.

Equipment

required

A 50ĂW SMAĂcoaxial cable and an oscilloscope (TDS820).

Prerequisites

The AWG610 Arbitrary Waveform Generator must meet the

prerequisites listed onpage B-8.

Do the following steps to install the test hookup and set the test equipment
controls:

1. Use the 50

W SMA coaxial cable to connect the AWG610 Arbitrary

Waveform Generator CH1 output connector to the oscilloscope CH1 input
connector (see Figure B–11).

ББ

ББ

AWG610 Arbitrary Waveform Generator

Oscilloscope (TDS820)

50 W SMAĂcoaxial cable

Figure B-11: Direct DA output pulse rise time initial test hookup

Check Pulse Rise Time

Advertising