Appendix B: Performance Verification

AWG610 Arbitrary Waveform Generator User Manual

B-49

Figure B-24: Waveform output when the SW3 is closed

c.

Open SW3 of the ground closure to degenerate the event signal.

d. Verify that the oscilloscope displays the waveform in Figure B–21.

10. Check the EVENT IN connector pin 3 input by doing the following

substeps:

a. Close the SW4 of the ground closure to generate an event signal on the

EVENT IN connector pin 3.

b. Verify that the oscilloscope displays the waveform shown in Figure

B–25.

Figure B-25: Waveform output when SW4 is closed