Tektronix AWG610 User Manual

Page 285

Advertising
background image

The UTILITY Window

AWG610 Arbitrary Waveform Generator User Manual

3-167

The manual diagnostics routines can execute a full set of hardware tests for all
the test categories or only for the specified category except for the DAC. You can
also specify a test cycle of 1 to infinite times.

Do the following steps to execute the diagnostics:

1. Push the RUN button to turn the output off if a waveform is being output.

The RUN LED turns off.

2. Push UTILITY (front-panel)

!Diag (bottom).

The screen shown in Figure 3–55 appears.

3. Push the Diagnostic xxxx side button and select a test category by using the

general purpose knob.

The xxxx represents currently selected test category. You can select a test
category from All, System, Run Mode, Clock, Output, Seq Mem and Wave
Mem. If you select All, the diagnostic routines of all categories are executed.

4. Push the Cycle n side button and select a test cycle by using the general

purpose knob.

The n represents a currently selected test cycle. You can select a test cycle
from 1, 3, 10, 100 or Infinite. If you select Infinite, the diagnostic tests are
repeated infinitely. Push the Abort Diagnostic side button to stop the
execution.

5. Push the Execute Diagnostic side button to start the diagnostic tests.

The - - - is displayed at each test category on the screen either at the beginning or
after the factory reset. The mark - - - is also displayed while the diagnostic test is
executing. See Figure 3–56. When the diagnostic test terminates without error,
Pass is displayed instead of the - - -. The test routine displays the error code and
skips to the next test if an error is detected.

See Table 3–45 for the test categories and error codes.

Table 3-45: Diagnostic categories and error codes

Categories

Error codes

Descriptions

System

2101 to 2106

Bios test error

2111 to 2116

FrontĆpanel test error

2301

A30 board test error

2401 to 2402

Clock delay data test error

2701 to 2702

Cal data test error

Manual Diagnostics

Advertising