Appendix B: Performance Verification

B-38

AWG610 Arbitrary Waveform Generator User Manual

Internal Trigger Tests

These procedures check internal trigger function of the AWG610 Arbitrary
Waveform Generator.

NOTE. When you output signal from the CH1 or CH1 OUTPUT, check that the
other OUTPUT ( CH1 or CH1 ) LED is off.

If the other OUTPUT LED is on, push the CH1 or CH1 OUT button to turn off
the output.

Equipment

required

A 50 W SMAĂcoaxial cable,ĂaĂSMA(Fe)ĆBNC(Ma)Ăadapter and an

oscilloscopeĂ(TDS700).

Prerequisites

The AWG610 Arbitrary Waveform Generator must meet the

prerequisites listed on page B-8.

Do the following steps to install the test hookup and set the test equipment
controls:

1. Use the coaxial cable and SMA(Fe)-BNC(Fe) adapter to connect the

AWG610 Arbitrary Waveform Generator CH1 output connector to the
oscilloscope CH1 input connector (see Figure B–14).

ББ

ББ

OscilloscopeĂ(TDS700)

50 W SMAĂcoaxial cable

AWG610 Arbitrary Waveform Generator

SMA(Fe)ĆBNC(Ma)

adapter

Figure B-14: Internal trigger initial test hookup

2. Set the oscilloscope controls as follows:

Vertical . . . . . . . . . . . . . . . . . . . . . . . . . CH1

CH1 coupling . . . . . . . . . . . . . . . . DC
CH1 scale. . . . . . . . . . . . . . . . . . 0.5 V/div
CH1 input impedance . . . . . . . . . . 50 W

Horizontal