Pulse response tests – Tektronix AWG610 User Manual

Page 394

Advertising
background image

Appendix B: Performance Verification

B-34

AWG610 Arbitrary Waveform Generator User Manual

Pulse Response Tests

This procedure checks the pulse response characteristics of the AWG610
Arbitrary Waveform Generator output waveforms at amplitudes of 1 V.

Equipment

required

A 50ĂW SMAĂcoaxial cable and an oscilloscope (TDS820).

Prerequisites

The AWG610 Arbitrary Waveform Generator must meet the

prerequisites listed on page B-8.

Do the following steps to install the test hookup and set the test equipment
controls:

1. Use the 50

W SMA coaxial cable to connect the AWG610 Arbitrary

Waveform Generator CH1 output connector to the oscilloscope CH1 input
connector (see Figure B–12).

ББ

ББ

AWG610 Arbitrary Waveform Generator

Oscilloscope (TDS820)

50 W SMAĂcoaxial cable

Figure B-12: Pulse response initial test hookup

2. Set the oscilloscope controls as follows:

Vertical . . . . . . . . . . . . . . . . . . . . . . . . . CH1

CH1 coupling . . . . . . . . . . . . . . . . DC if applicable
CH1 scale . . . . . . . . . . . . . . . . . . 250 mV/div (200 mV for TDS820)

Horizontal

Sweep . . . . . . . . . . . . . . . . . . . . . 500 ps/div

Trigger

Source . . . . . . . . . . . . . . . . . . . . . CH1
Slope . . . . . . . . . . . . . . . . . . . . . . Positive
Level . . . . . . . . . . . . . . . . . . . . . . 0 V
Mode . . . . . . . . . . . . . . . . . . . . . . Auto

Advertising