Appendix B: Performance Verification

AWG610 Arbitrary Waveform Generator User Manual

B-53

b. Set the function generator (AFG310) controls:

Function . . . . . . . . . . . . . . . . . . . . . . . . Square
Mode . . . . . . . . . . . . . . . . . . . . . . . . . . Continuous
Parameters . . . . . . . . . . . . . . . . . . . . . .

Frequency . . . . . . . . . . . . . . . . . . 10 MHz
Amplitude . . . . . . . . . . . . . . . . . . . 2.0 V into 50 W (4.0 V into 1 MW)
Offs et . . . . . . . . . . . . . . . . . . . . . . 0 V

Output . . . . . . . . . . . . . . . . . . . . . . . . . On

1. Follow the substeps below to set the AWG610 Arbitrary Waveform

Generator controls and select the waveform file:

a. Push UTILITY (front-panel)

!System (bottom)!Factory Reset

(side)

!OK (side).

b. Load the MODE.WFM file.

Refer to Loading Files on page B–10 for file loading procedures.

c.

Push HORIZONTAL MENU (front-panel)

!Clock (side).

d. Push 2, 0, 0 and M (SHIFT+7) keys in this order or turn the general pur-

pose knob to set the internal clock frequency to 200 MHz.

2. Push the RUN button.

The LED above the RUN button is on.

3. Verify that the frequency counter reading is 50 MHz

±

50 Hz.

4. Retain the test hookup.

5. Push HORIZONTAL MENU (front-panel)

!Clock Ref (side) so that the

AWG610 Arbitrary Waveform Generator clock reference is set to External.

6. Verify that the frequency counter reading is 50.0 MHz

±

10 kHz (using exter-

nal reference clock).

7. Modify the function generator controls as follows:

Frequency . . . . . . . . . . . . . . . . . . . . . . 10.1 MHz

8. Check that the frequency counter reading is 50.5 MHz

±

10 kHz.

9. Turn the function generator output off and disconnect the function generator

and frequency counter.

CheckĂ1/4ĂClock

frequency

CheckĂ10MHzĂReference

InputĂ