Operating mode tests – Tektronix AWG610 User Manual

Page 377

Advertising
background image

Appendix B: Performance Verification

AWG610 Arbitrary Waveform Generator User Manual

B-17

Operating Mode Tests

The following procedures verify the operation of the Cont, Triggered and Gated
modes.

NOTE. When you output signal from the CH1 or CH1 OUTPUT, check that the
other OUTPUT ( CH1 or CH1 ) LED is off.

If the other OUTPUT LED is on, push the CH1 or CH1 OUT button to turn off
the output.

Equipment

required

A 50 W SMAĂcoaxial cable,ĂaĂSMA(Fe)ĆBNC(Ma)Ăadapter and an

oscilloscopeĂ(TDS700).

Prerequisites

The AWG Arbitrary Waveform Generator must meet the prerequisites

listedon page B-8.

Do the following steps to install the test hookup and set the test equipment
controls:

1. Use a 50

W

coaxial cable to

c

onnect the AWG610 Arbitrary Waveform

Generator CH1 output connector to the oscilloscope CH1 input connector
(see Figure B–5).

OscilloscopeĂ(TDS700)

50 W SMAĂcoaxial cable

AWG610 Arbitrary Waveform Generator

SMA(Fe)ĆBNC(Ma)

adapter

ББ

ББ

Figure B-5: Cont mode initial test hookup

2. Set the oscilloscope controls as follows:

Vertical . . . . . . . . . . . . . . . . . . . . . . . . . CH1

CH1 coupling . . . . . . . . . . . . . . . . DC
CH1 scale . . . . . . . . . . . . . . . . . . 0.2 V/d iv
CH1 input impedance . . . . . . . . . . 50 W

Horizontal

Sweep . . . . . . . . . . . . . . . . . . . . . 200 ns/div

Check Cont Mode