Xylem photoLab 6600 UV-VIS User Manual
Page 101

photoLab
®
6600 UV-VIS
Operation
101
ba75848e01
08/2009
1
Select the required method in the
Concentration
mode.
Manual selection of the method
(see section 4.5.5).
2
Open the Timer menu.
The Timer menu is open.
3
If necessary, add a new timer to
the list with [Add].
Note:
The [Add] function key is only
displayed if a method is selected
for which analysis timers were
programmed but are not yet
displayed in the list of timers.
4
Highlight an analysis timer.
5
If necessary, remove the analysis
timer from the list with [Remove].
6
Start the highlighted timer with
[Start].
The status of the timer is Active.
When the specified time interval
has expired, an acoustic signal
sounds and the status changes to
Expired.
7
Stop the highlighted timer with
[Stop].
The status of the timer changes to
Inactive. The acoustic signal is
switched off.
<TIMER>
Timer
04/16/07 9:52
Designation
Time
Status
User defined timer
00:15:00
Inactiv
A6/25 - 1
00:15:00
Inactiv
Start
Stop
Remove
Add