Xylem photoLab 6600 UV-VIS User Manual

Page 63

Advertising
background image

photoLab

®

6600 UV-VIS

Operation

63

ba75848e01

08/2009

How to program

user-defined

methods

Note
During the following proceeding, you can return to the previous page at any
time with [Back], e. g. if you want to correct entries, add further value pairs or
eliminate outliers.

1

Enter the general method data
here. The next available method
number is already entered as the
number.

You have the following options
when filling out the input fields:

- Fill out all empty input fields one

after the other

- Using [Method list], select an

already existing method as a
model, give it a new method
number and adjust the entries

- Using [Method list], select an

existing method in order to
change it (without changing the
number).

- You can delete the method

completely with [Delete].

2

Select the menu item, Calibration
curve
. Select the method for the
determination of the calibration
line. The following variants can be
selected:

- Measure standard solutions

- Enter value pairs

- Enter formula

3

Using [Next], accept all entries on
the page and switch to the next
page.

<HOME>
Concentration

[Setup]

New method

Edit method

04/16/07 9:52

Number

1001

Designation

Nitrite

Version

01

Wavelength

525

Cell

10 mm

Citation form

NO2-N

Unit

mg/l

Resolution

0.001

Calibration curve

Measure standard solutions

Method list

Delete

Next