Xylem photoLab 6600 UV-VIS User Manual

Page 38

Advertising
background image

Operation

photoLab

®

6600 UV-VIS

38

ba75848e01

08/2009

Note
Ordering information is given in chapter 8 A

CCESSORIES

AND

OPTIONS

. The

cells listed in the chapter 8 A

CCESSORIES

AND

OPTIONS

are especially adapted

to the WTW test set program. General requirements of the cells are given in
chapter 7 T

ECHNICAL

DATA

. Note that the spectral transparency of the cell

must be suitable for the intended application (example, quartz cell for UV
range).

Carrying out a zero

adjustment

The zero adjustment takes place similarly in the Concentration, Absorbance
/ % Transmission
, Multi wavelengths and Kinetics modes.

1

In the respective mode, press the
<ZERO·BLANK> key.

2

In Concentration mode only:

Select and confirm Zero
adjustment
.

The zero adjustment window pops
up.

3

Close the inner turn-up lid.

4

Depending on the cell type, insert
the zero cell as follows:

Round cell:

Insert the round cell in the round
cell shaft so it touches the bottom.

If the inner turn-up lid is opened
too wide, a message prompts you
to close the inner turn-up lid.

Concentration

04/16/07 9:52

3: A6/25

NH

4

-N

16 mm

0.20 - 8.00 mg/l

Setup

Method list

Citation form

Unit

Adjust

Blank value

Zero adjustment

Zero adjustment

04/16/07 9:52

Please insert zero cell (distilled water)

or press <Start/Enter>

Zero cell
(H O dist.)

2

inner
turn-up lid