Operation

photoLab

®

6600 UV-VIS

122

ba75848e01

08/2009

Activating AQA1

The AQA1 monitoring is activated in the Mode menu:

Defining the AQA1

Interval

The AQA1 Interval defines the interval between two AQA1 checks.
When an interval has expired, the following consequences become effective:

z Warning and loss of the AQA1 labeling

z Locking of the photometer against all measurements (if activated).

Select and confirm Weeks.

AQA1 is active.
The Interval setting indicates
Weeks as the interval unit.

<HOME>
[AQA]

AQA1 setup

Mode

AQA1 setup

04/16/07 9:52

Mode

Weeks

Interval
Lock instrument
Configuration of tests ...

Mode

AQA1 inactive

Weeks

1

Enter a numeric value (2 to 52
weeks) (<0...9>) and confirm

The Interval defined for the AQA1
check is active.

<HOME>
[AQA]

AQA1 setup

Interval

AQA1 setup

04/16/07 9:52

Mode

Weeks

Interval

Lock instrument
Configuration of tests ...

Interval

2

Weeks