Carrying out the

AQA1 check

(example:

PhotoCheck)

The AQA1 check comprises the check with all test standards activated in the
menu, AQA menu / AQA1 setup / Configuration of tests ... for AQA1 (see
page 123).

First, a zero adjustment for all wavelengths takes place.
Subsequently, the first individual checks with the selected test standards take
place (e.g. PhotoCheck).

Example, PhotoCheck:

5

Using <

><> and <W><X>,

select the Lot number, Target
value
or Tolerance entries and
open them for editing with
<START·ENTER>

.

6

Enter and confirm the required
value (<0...9>)

7

Accept all values with [Apply].

PhotoCheck

04/16/07 9:52

Lot number:

HC616115

Use by

16.04.2008

Target value

Tolerance

445/1

0.196

± 0.020

445/2

0.500

± 0.030

445/3

0.998

± 0.040

445/4

1.508

± 0.050

525/1

0.197

± 0.020

525/2

0.495

± 0.030

525/3

0.992

± 0.040

525/4

1.496

± 0.050

Apply

The photometer is ready for the
zero adjustment.

1

Insert the zero cell.
The cell is automatically
recognized and the zero
adjustment is started for all
wavelengths.

After the successful zero
adjustment, the photometer is
ready to measure the PhotoCheck
test standard
445/1.

<HOME>
[AQA]

AQA1 check

PhotoCheck

04/16/07 9:52

Reference measurement

Please insert zero cell (distilled water).