The photometer is ready to
measure.

1

Open the setting menu with
[Setup].

2

Select and confirm Sample blank
value
.

3

Insert the cell with a suitable blank
sample.

The first single measurement for
the sample blank value takes
place.

The following data is displayed as
the result:

– The measured absorbance

from the (last) single
measurement.

– The median from all single

measurements carried out up to
now.

4

If necessary, carry out further
single measurements for the
formation of the median with [Next
meas.]
or
discard the last single
measurement with [Discard].

5

To accept the median value, press
[Apply].

The photometer is ready to
measure.

The use of the sample blank value
is indicated by [SB] in the top right
corner of the display.

Concentration

04/16/07 9:52

To start measurement,

insert cell or press <Start/Enter>

3: A6/25

NH

4

-N

16 mm

0.20 - 8.00 mg/l

Setup

Method list

Citation form

Unit

Sample blank value

04/16/07 9:52

To start measurement,

insert cell or press <Start/Enter>

3: A6/25

NH

4

-N

16 mm

0.20 - 8.00 mg/l

Sample blank value

04/16/07 9:52

Last measured absorbance

0.115

Median

0.115 (1 Measurement(s))

3: A6/25

NH

4

-N

16 mm

0.20 - 8.00 mg/l

Next meas.

Discard

Apply

Concentration

04/16/07 9:52

[SB]

To start measurement,

insert cell or press <Start/Enter>

3: A6/25

NH

4

-N

16 mm

0.20 - 8.00 mg/l

Setup

Method list

Citation form

Unit