Variant 2:

Enter value pairs

Unlike variant 1, the fields of the Absorbance column are filled out manually
here. Accordingly, the steps 6 to 10 are not applicable here. Apart from that,
the proceeding is identical to variant 1.

The related formula f(x) and
correlation coefficient R

2

are

displayed above the diagram.

12 If required, select a different line

type for the line adjustment with
[Curve type].

- Polygon line

- Straight line

- Parabola

13 If required, enter different

measured value limits with [Meas.
range]
.

- Lower limit

- Upper limit

14 Using [Next], complete the editing

of the calibration line and proceed
to the next page.

The timers and AQA2 data linked
to the method are displayed.

15 If necessary, enter intervals for up

to 4 timers.

16 If necessary, enter the AQA2

target value and AQA2 tolerance.

17 Complete the programming of the

method with [Complete].

The method is programmed and
selected for measuring.

Edit method

04/16/07 9:52

f(x)=0.73x
R

2

= 1.000

Meas. range: 0.050 - 1.000 mg/l

Concentration [mg/l NO2-N]

Back

Curve type

Meas. range

Next

Ab

so

rb

anc

e

Edit method

04/16/07 9:52

Timer 0

00:00:00

Timer 1

00:00:00

Timer 2

00:00:00

Timer 3

00:00:00

AQA2 target value

1.00 mg/l

AQA2 tolerance

0.10 mg/l

Back

Complete