6 what to do if, 1 actions in the case of a broken cell, What to do if – Xylem photoLab 6600 UV-VIS User Manual

Page 153: Actions in the case of a broken cell, 6what to do if

Advertising
background image

photoLab

®

6600 UV-VIS

What to do if ...

153

ba75848e01

08/2009

6

What to do if ...

6.1

Actions in the case of a broken cell

Warning

Cells can contain dangerous substances. If the contents are released,
follow the safety instructions of the package insert. If necessary, take
corresponding protective measures (protective goggles, protective
gloves etc.).

Caution

Do not turn the photometer upside down to remove the liquid!
When doing so, the liquid could come into contact with electronic
components and damage the photometer.

The photometer has a drain device through which the contents of a broken
cell can drain off without causing any damage.

Proceeding after a

cell has broken

Note
After recommissioning, carry out the photometer monitoring for all
measurements (see section 4.14.2).

1

Switch off the photometer and
disconnect it from the power
supply.

2

Let the liquid drain off into a
suitable container and dispose of
it properly according to the
instructions of the reagent
package

3

Carefully clean the cell shaft using
a moist, lint-free cloth

4

Let the cell shaft dry.