4.14.2 Photometer monitoring (AQA1)

At least one set of test standards such as Spectroquant

®

PhotoCheck or

CertiPUR

®

is required for the photometer monitoring.

The administrator specifies which test standard has to be used as the
minimum requirement for the AQA1 monitoring.
The extent of the monitoring can be enlarged with further test standards.

Note
Settings for AQA checks are only available for users of the user group,
administrator.
Every registered user can carry out the AQA check (see also section 4.15.1).

Spectroquant

®

PhotoCheck

The PhotoCheck consists of 12 test standards in duplicate, 2 zero cells and
2 cells to check the barcode reader. Each PhotoCheck package contains a
lot dependent test certificate with all nominal values (absorbances) and
tolerances of the test standards. These values are entered in the photometer
during the configuration of the AQA1 check.

CertiPUR

®

test

standards

Each CertiPUR

®

standard is provided with a lot dependent test certificate

with all nominal values (absorbances) and tolerances of the test standards.
The values were preset in the factory.

Note
Observe the shelf life of the test standards. The values in the photometer
always have to be checked when a new package of test standard is used. If
necessary, adjust the values at the photometer.

Overview of the

photometer

monitoring

Photometer monitoring (AQA1) consists of the following parts:

z Configuring settings in the AQA1 setup menu.

– Activate AQA1

– Specify AQA1 Interval

– Activate/deactivate the meter lock for missing or expired AQA1 check

– Define the extent of the AQA1 monitoring by activating or deactivating

the individual test standards.

– Enter the nominal values, tolerances and lot numbers for the individual

test standards

z Carrying out the AQA1 check. The photometer compares the results with

the nominal values while taking into account the tolerances.

The steps are described in detail below.