Operation

photoLab

®

6600 UV-VIS

134

ba75848e01

08/2009

The measurement display
appears.

12 Insert the cell with the respective

sample.

The sample is measured.

After the measurement, the
recovery rate is displayed in the
right table column.

The assessment of the recovery
rate is displayed next to
the recovery rate
(

✓ or ✗

).

The criteria for the assessment
are determined in the menu,
AQA3/MatrixCheck setup /
Maximum difference.

13 If necessary, repeat the steps 11

and 12 for the remaining samples.

14 Use [Complete] to complete the

MatrixCheck.

The Save dialog box pops up.

15 If necessary, change the storage

location with [Location]:
Internal DataB folder:
Exchange folder in the instrument
or
USB memory:
USB memory device connected at
the USB-A connection.

16 If necessary, change the file

name.

17 Save the file with

<START·ENTER>.

MatrixCheck

04/16/07 9:52

Method

1: C3/25

Sample concentration

45 mg/lCOD

Sample

10 ml

Standard

0.5 ml

To start measurement,

insert cell or press <Start/Enter>

16 mm

Back

MatrixCheck

04/16/07 9:52

Method

1: C3/25

Sample concentration

45 mg/lCOD

Sample

Standard

Target value

nominal

[ml]

[ml]

[mg/l]

[mg/l]

10

0.5

62

58

94 %

10

1

77

10

1.5

91

Back