Xylem photoLab 6600 UV-VIS User Manual
Page 135

photoLab
®
6600 UV-VIS
Operation
135
ba75848e01
08/2009
Test record
The result of the MatrixCheck is displayed in a test record. You can print this
record and save it as a file.
To save the file in the photometer, select the Internal DataB folder as the
location. To save the file in an external USB memory device at the USB-A
connection, select USB memory as the location (see section 4.11.1).
Sample printout:
The display returns to the
measured value display of the
original sample without spiking /
dilution.
The [MC] status indicator is
displayed. A MatrixCheck was
carried out for this measured
value.
Concentration
04/16/07 9:52
[MC]
45
mg/l
18: C3/25
COD
16 mm
10 - 150 mg/l
Setup
Method list
Citation form
Unit
SpectroFlex 6600 09130512 1.30-WTW-1.60 Administrator
MatrixCheck
OK
Protocol ID
7
Method
1: C3/25 COD
Sample concentration
45
mg/lCOD
Standard ID
COD 1500
Standard concentration400
mg/lCOD
Sample
Standard
Target value
Actual value
ml
ml
mg/l
mg/l
10
0.5
62
58
94% OK
10
1
77
71
92% OK
10
1.5
91
77
85% OK