Validity of the zero

adjustment

The data of the zero adjustment is stored in the photometer separately for
each cell type. As long as the data is valid, it is automatically used again after
a temporary change to a different cell type. The validity depends on the
respective mode:

* After the wavelength or method respectively was temporarily exited the photometer

displays that a zero adjustment is available and the time it was carried out. You can then
decide whether to use this zero adjustment or carry out a new zero adjustment.

When to repeat the

zero adjustment?

We recommend to repeat the zero adjustment in the following cases:

z If the photometer was subject to mechanical stress such as strong shock

or transport

z If the ambient temperature changed by more than 5 °C since the last zero

adjustment

z At least once per week

z If a new cell type (different manufacturer, different glass type is used)

z Basically each time you want to measure with the highest possible

accuracy.

Mode

Validity of the zero adjustment

Concentration (permanently
programmed methods)

z Till the next zero adjustment

Absorbance / % Transmission

z Till the next zero adjustment with the

same wavelength *

Concentration (user-defined
methods) and

Multi wavelengths

z Till the next zero adjustment for the same

method *

Kinetics

z Till another kinetic profile is loaded
z Till the Kinetics mode is exited or the

photometer is switched off