Xylem photoLab 6600 UV-VIS User Manual

Page 72

Advertising
background image

Operation

photoLab

®

6600 UV-VIS

72

ba75848e01

08/2009

2

Insert the cell with the reference
sample.

The first single measurement for
the Reference absorbance is
carried out.

The following data is displayed as
the result:

– The measured absorbance

from the (last) single
measurement.

– The median from all single

measurements carried out up to
now.

3

If necessary, carry out further
single measurements for the
formation of the median with [Next
meas.]
or
discard the last single
measurement with [Discard].

4

To accept the median value, press
[Apply].

The photometer is ready to
measure.

The reference absorbance is
displayed in the top right corner
during absorbance or
transmission measurement.

Reference absorbance

04/16/07 9:52

To start measurement,

insert cell or press <Start/Enter>

489 nm

10 mm

Reference absorbance

04/16/07 9:52

Last measured absorbance

0.232

Median

0.232 (1 Measurement(s))

489 nm

10 mm

Next meas.

Discard

Apply

Absorbance

04/16/07 9:52

Reference::

To start measurement,

insert cell or press <Start/Enter>

489 nm

10 mm

Setup

Wavelength

Transmission

Reference