Xylem photoLab 6600 UV-VIS User Manual

Page 127

Advertising
background image

photoLab

®

6600 UV-VIS

Operation

127

ba75848e01

08/2009

General AQA2

settings

Note
When the mode (Weeks or Measurements) is changed, all AQA2 intervals
are reset to the preset values.

Locking the method

Here you configure whether or not a method will be locked against
measurement if there is no valid AQA2 check or the interval for the AQA2
check has expired.

1

Select and confirm Mode.
The Mode selection field pops up.

2

Select and confirm Weeks or
Measurements.

AQA2 is active. For all methods,
the AQA2 intervals are entered
either in weeks or number of
measurements.

3

Accept the general settings with
[Apply].

<HOME>
[AQA]

AQA2 setup

AQA2 setup

04/16/07 9:52

Mode

Weeks

Lock methods

Yes

Method...

Method list

1

Select and confirm Lock methods.

2

Select and confirm Yes.

The method lock is enabled.

Each method will be locked if the
AQA2 check is invalid or the AQA2
interval has expired.

<HOME>
[AQA]

AQA2 setup

AQA2 setup

04/16/07 9:52

Mode

Weeks

Lock methods

Yes

Method...

Method list