Operation

129

ba75848e01

08/2009

Carrying out the

AQA2 check for a

method

Test record

A test record is displayed after the check. It can be printed and stored as a
file (in the internal DataB folder or USB memory device at the USB-A
connection, see section 4.11.1).

Sample printout:

Note
Later you can view the last AQA2 test records for all methods monitored with
AQA2 under AQA2 info.

1

Carry out the check like a normal
measurement (see section 4.5.1
to 4.5.3).

2

Insert the cell or
start measurement with
<START·ENTER>.

After the measurement is
completed, the result and its
evaluation are displayed.

If the check failed, it is possible to
repeat the measurement.

If the check was successful, the
AQA2 check function is finished.

<HOME>
[AQA]

AQA2 check

AQA2 check

04/16/07 9:52

Target value

2.00

To start measurement,

insert cell or press <Start/Enter>

3: A6/25

NH

4

-N

16 mm

0.20 - 8.00 mg/l

SpectroFlex 6600 09130512 1.30-WTW-1.60 Administrator

AQA2

OK

Protocol ID

32

Executed by:

Administrator

Executed

21.05.2007

Valid until:

13.08.2007

Method

6: P6/25 PO4-P

Standard ID

CC10 OC557775

Target value

0.80 +- 0.08 mg/l

Measured value

0.84

mg/l