Operation

photoLab

®

6600 UV-VIS

56

ba75848e01

08/2009

Measuring the

reagent blank value

Inserting a cell with barcode starts
a measurement.

If a cell without barcode is used:
Select the method manually
(see section 4.5.5).

The photometer is ready to
measure.

1

Using <ZERO·BLANK>, open the
Adjust selection list.

2

Select and confirm Blank value.

The window for the measurement
of the reagent blank value pops
up.
The data of the last measurement
appears in the measured value
display.

<HOME>
Concentration

Concentration

04/16/07 9:52

Please select method for measuring
or insert a barcoded cell
or insert AutoSelector.

Setup

Method list

Last method

New Method

Concentration

04/16/07 9:52

To start measurement,

insert cell or press <Start/Enter>

3: A6/25

NH

4

-N

16 mm

0.20 - 8.00 mg/l

Setup

Method list

Citation form

Unit

Concentration

04/16/07 9:52

3: A6/25

NH

4

-N

16 mm

0.20 - 8.00 mg/l

Setup

Method list

Citation form

Unit

Adjust

Zero adjustment

Blank value