Operation

photoLab

®

6600 UV-VIS

64

ba75848e01

08/2009

Variant 1:

Measure standard

solutions

1

Select and confirm Measure
standard solutions
.

2

Enter and confirm details of the
standard solutions (optional).

3

Using [Next], accept all entries on
the page and switch to the next
page.

The table for the measurement of
standard solutions pops up.

In the first two lines of the table,
the two value pairs (measuring
points) that are at least required
for a calibration are already
prepared (reagent blank value E0
and any further nominal value).

4

Create further values pairs with
[Add] as necessary.

You can delete a highlighted value
pair with [Delete].

5

In the Target value column, enter
the nominal values of the
individual standard solutions.

Measuring the standard solutions:

6

Using the arrow keys <

><>

and <

W><X>, navigate to the

relevant input field in the
Absorbance column and press
<START·ENTER>.

Edit method

04/16/07 9:52

Standard ID

Standard manufacturer

Back

Next

Edit method

04/16/07 9:52

Target value

Absorbance

E0

0.000

1

Back

Add

Delete

Next

Edit method

04/16/07 9:52

Target value

Absorbance

E0

0.000

1

0.300

2

0.600

3

1.000

Back

Add

Delete

Next

Edit method

04/16/07 9:52

Target value

Absorbance

E0

0.000

1

0.300

2

0.600

3

1.000

Back

Add

Delete

Next