Home> [aqa] – aqa1 setup – configuration of tests – Xylem photoLab 6600 UV-VIS User Manual

Page 123

Advertising
background image

photoLab

®

6600 UV-VIS

Operation

123

ba75848e01

08/2009

Configuring the lock

of the photometer

Here you configure whether or not the photometer will be locked against all
measurements if there is no valid AQA1 check or the interval for the AQA1
check has expired.

Configuration of

tests ...

1

Select and confirm Yes.

The photometer is locked against
all measurements if the AQA1
check is invalid or the AQA1
interval has expired.

<HOME>
[AQA]

AQA1 setup

Lock instrument

AQA1 setup

04/16/07 9:52

Mode

Weeks

Interval

Lock instrument

Configuration of tests ...

Lock instrument

Should the instrument be locked for further

measurements if AQA1 check is invalid or

has expired?

No

Yes

All possible test standards or test
standard sets are listed.

1

Select and confirm a test standard
or test standard set.

2

Adjust and confirm the extent of
the monitoring with Activate or
Deactivate.

3

Confirm the test standard (set)
once again.

4

Switch to the adjustment of the
nominal values and tolerances
with Setup

.

<HOME>
[AQA]

AQA1 setup

Configuration of tests ...

AQA1 setup

04/16/07 9:52

PhotoCheck

Active

CertiPUR UV-VIS 1

Inactive

CertiPUR UV-VIS 1A

Inactive

CertiPUR UV-VIS 2

Inactive

CertiPUR UV-VIS 3

Inactive

CertiPUR UV-VIS

Inactive

I

Apply

CertiPUR UV-VIS 1
General setup

Activate