4.13.3 Printing measurement datasets

This section describes how to print measurement datasets of the measuring
modes, Concentration, Absorbance / % Transmission, and Multi
wavelengths
.

By means of sample printouts, the printed information is described below:

Concentration

and Multi

wavelengths mode

Structure of the lines from left to right:

1st line:

[Consecutive no.] [Date] [Time] [Method name] [Measured value] [Unit]
[Citation form] [Dilution] [ID or "AutoStore"]

2nd line (long version only):

[User] [Reagent blank value] [Date of blank value measurement]
[Time of blank value measurement] [Lot ID of blank value measurement]
[AQA1:
label] [AQA1: record no.] [AQA2: label] [AQA2: record no]

Note
Optional elements (e.g. dilution or ID) are output only if they were really used
for measurement or storage.

Absorbance / %

Transmission

mode

Structure of the lines from left to right:

1st line:

[Consecutive no.] [Date] [Time] [Wavelength] [Measured value]
["Absorbance" or "Transmission" mode ] [ID or "AutoStore"]

2nd line (long version only):

[User] [Value of reference absorbance] [Date of reference absorbance] [Time
of reference absorbance] [AQA1:
label] [AQA1: record no.]

Note
Optional elements (e.g. ID or reference absorbance) are output only if they
were really used for measurement or storage.

21 05.06.07 14:05:41 C4/25 844 mg/l COD

Inlet

Administrator 0.005 02.06.07 11:02:13 2 AQA1: 9 AQA2: 14

14 05.06.07 11:25:01 445 nm 0.609 Absorbance AutoStore

Administrator 0.133 02.06.07 09:59:01 AQA1: 9