Self test

During the self-test, all cells must be removed and the cell shaft cover closed.

Note
The result of the self-test can be viewed and printed with the [Info] function
key (see section 4.17).

Warm-up time

After switching on the photometer requires a warm-up time of 15 minutes.
Reproducibility of measurement data is restricted during the warm-up time.
Therefore, do not measure during the warm-up time.

During the warm-up time, a progress bar appears on the display next to the
date. The progress bar disappears as soon as the warm-up time is over.

The self-test includes:

– the test of the memory,

processor,
internal interfaces,
filter and lamp

– a calibration for each

wavelength

After the self-test is completed,
the main menu is displayed.

Self test

04/16/07 9:52

Keep cover closed

System test
Filter test
Lamp test
Wavelength calibration

Progress bar during warm-up
time