Advanced operations – Hach-Lange DR 2800 eco User Manual

Page 101

Advertising
background image

99

Advanced Operations

6.5.3.1 Navigating in the graph of a time scan or a time scan analysis

After a time scan has been carried out, the time and the
absorbance/transmittance data are displayed as a curve.

Where the cursor is positioned on the curve, the elapsed time up to
this point and the corresponding absorbance are highlighted.

Table 15 Navigation within the "time scan” curve

Cursor Function/
Zoom Function

Description

Curve Icon
(Choice of Cursor Mode)

Delta mode: A second cursor is activated. The position of the fixed cursor was previously
defined in Cursor Mode Single. You can use the active cursor to select any point on the
measurement curve. The difference to the fixed cursor is shown on the curve. The delta values
are correspondingly highlighted and displayed on the x and y axes.
The gradient of the curve and the correlation coefficient (r

2

) between the cursor points in the

Delta mode are shown under the curve.

Cursor Mode Single: The cursor moves to each selected measurement point of the scan.

Arrow keys

The arrow keys (right/left) are used to move the cursor (depending on the selected mode) to the
next data point. The data of the data point (wavelength/absorbance or transmittance value) are
highlighted on the x and y axes.

Note: You can also press any point on the curve to display the associated data.

Zoom Icon

This function is used to magnify the section of the curve in the vicinity of the cursor. The original
curve size can be restored by pressing the zoom icon again.

Advertising
This manual is related to the following products: