1 carrying out a standard addition, Advanced operations – Hach-Lange DR 2800 eco User Manual

Page 87

Advertising
background image

85

Advanced Operations

g. Is the pipette volume correct?

2.

Check the reagents you used, by repeating the standard
addition procedure with freshly prepared reagents.

If none of these measures resolves the problem, please contact the
manufacturer or a sales representative.

6.2.1 Carrying out a standard addition

Carry out the standard addition in accordance with the
corresponding HACH procedure.

There are two different methods:

Peak volume (Standard addition):
Defined volumes of a standard solution are added step by step to
an already analyzed sample. The sample is measured again after
each addition.

Sample volume:
A defined volume of a standard solution of known concentration is
added to the volume of sample specified in the procedure and the
sample is measured after each addition. In most cases three
different standard solutions are prepared and the procedure is
repeated for each of them.

Note: The units and chemical forms used for the sample are used for the
standard solutions. Be careful to ensure that you use the correct units for
subsequent entries.

Peak volume/sample volume methods

1.

Select Stored Programs in the main menu. Select the required
program.

2.

Press Start.

3.

Analyze a sample without added standard solution in
accordance with the instructions in the Procedures Manual.
When the measurement is complete, leave the sample cell in
the cell holder.

Detection rate

Conclusion

100%

The originally used reagents were defective.

Check the standard solution:

Repeat the standard addition procedure with a
freshly prepared standard solution.

Detection rate

Conclusion

100%

The originally used
standard solution was
defective.