2 programming of a new lange cuvette test, Standard operations – Hach-Lange DR 2800 eco User Manual

Page 53

Advertising
background image

51

Standard Operations

1.

Press Stored Programs in the ”Main Menu“ and select a
program.

2.

Press Program Options and then Edit.

3.

The Data list of the corresponding test will appear.

The working procedures of the corresponding LANGE Cuvette
Test contains the new datas (Wavelength, Factors,
Measurement Ranges, Conversion Factor for chemical Form
a.s.o.).

4.

Activate the corresponding line in the data overview and press
Edit.

5.

Press OK and then Store when you have finished editing the
test.

5.4.8.2 Programming of a new LANGE Cuvette Test

1.

Press Stored Programs in the ”Main Menu“.

2.

Press Program Options and then New.

3.

Use the alphanumeric keypad to enter the Program Number
indicated in the working procedure.

A data overview, which should show the program points of the
test that is to be programmed, is displayed.

The working procedures of the corresponding LANGE Cuvette
Test contains the new datas (Wavelength, Factors,
Measurement Ranges, Conversion Factor for chemical Form
a.s.o.).

4.

Activate the line containing the test specification that is to be
programmed, then press Edit and enter the data in the
appropriate program lines.

5.

Press OK and then Store when you have finished editing the
test.