Maintenance – Hach-Lange DR 2800 eco User Manual

Page 114

Advertising
background image

112

Maintenance

CAUTION
Wait until the lamp cools down. Contact with the hot lamp can
cause burns.

9.

Lift the lamp fitting carefully.

10. Carefully remove both knurled screws and put them to one

side.

11. Remove the halogen lamp, complete with the plug panel.

12. Carefully unplug the halogen lamp from the plug panel.

Important Note: Hold the lamp by the fitting only. Avoid touching
the glass, as substances on the skin can bake onto the lamp bulb
and thus accelerate the ageing process of the lamp.

13. Plug new halogen lamp (order number LZV565) to the panel

again.

14. Insert the halogen lamp again with the rounded part pointing

down.

15. Secure firmly with the two knurled screws.

16. Carefully insert the lamp fitting again.

17. Use a screwdriver (slotted or cross-headed) to replace the

screws securing the lamp fitting.

18. Put the battery back in place, see

7.2 Inserting or changing the

battery (DR 2800 only) on page 107

.

19. Use a screwdriver or a coin to replace the screw securing the

cover.

20. Carefully stand the instrument upright.

21. Plug in power supply again – the Instrument is now ready for

use.

22. Reset the Lamp History, see

6.6.5 Lamp History on page 102

.