Standard Operations

5.4 Stored Programs (DR 2800 only)

The DR 2800 spectrophotometer contains more than 200
programmed procedures that you can access through the
Stored Programs menu.

5.4.1 Selecting a saved test/method; entering user-specific basic data

1.

Press Stored Programs in the ”Main Menu“ to see an
alphabetical list of stored programs with program numbers.

The Stored Programs list will appear.

2.

Highlight your selection by pressing it.

Note: Use the scroll bar to scroll through the list quickly.

Note: If you already know the number of the test or method, press
Select by Number. Use the alphanumeric keypad to enter the test
number and confirm your input by pressing OK.

3.

Press Start to run the program.

Note: After a program is selected, the screen for that parameter will
appear. You do not need to select the wavelength.

Note: Follow the chemical procedures described in the corresponding
Procedures Manual.

4.

Press Options for Parameter Setup.

Table 5 ”Stored Programs“ options

Figure 11 Options

Options

Description

More...

For further Options

Store Off/On

With the Store On setting, all measurement data are stored automatically. With the Store Off
setting, no measurement data are stored.

% Trans/Conc/Abs

To switch to % transmittance, concentration or absorbance readings