Standard operations – Hach-Lange DR 2800 eco User Manual

Page 63

Advertising
background image

61

Standard Operations

5.5.6.4 Programming of a new LANGE Cuvette Test (DR 2800 only)

1.

Insert the Light Shield.

2.

The instrument need to be at the ”Main Menu“ or press
Barcode Programs in the ”Main Menu“.

3.

Insert sample cuvette or blank (depending on the working
procedure) into the cell compartment.

4.

After the comment "Program not available", a data overview is
displayed, including the specifications of the test that is to be
revised.

The working procedures of the corresponding LANGE Cuvette
Test contains the new data (Wavelength, Factors,
Measurement Ranges, Conversion Factor for chemical Form
a.s.o.).

5.

Activate the corresponding line including the specifications of
the test that is to be programmed and press Edit.

6.

When you have finished programming, press Store and then
Cancel to return to the ”Main Menu“.

7.

Touch Start to start the Cuvette Test with the new data.