5.5.6.3 Updating an existing Barcode Cuvette Test

Note: Only in very few cases does the revision of a test require all test
specifications to be updated.

Option 1 Manual Update of Test Data

1.

Insert the Light Shield.

2.

The instrument need to be at the ”Main Menu“ or press
Barcode Programs in the ”Main Menu“.

3.

Insert sample cuvette or blank (depending on the working
procedure) into the cell compartment.

4.

After the comment "Barcode controlnumber?", a data overview
is displayed, including the specifications of the test that is to be
revised.

Note: The working procedures of the corresponding LANGE Cuvette
Test contains the new data (Wavelength, Factors, Measurement
Ranges, Conversion Factor for chemical Form a.s.o.).

5.

Activate the corresponding line in the data overview and press
Edit.

6.

When you have finished programming, press Store and then
Cancel to return to the ”Main Menu“ menu.

7.

Touch Start to start the Cuvette Test with the new data.

Option 2 Manual Check/Revision of Test Data
If an incorrect entry was made during the data update and is not
recognized via the barcode of the cuvette (e.g. an incorrect
measuring range), another option is available for checking, and if
necessary revising, the test data.

1.

Insert the Light Shield.

2.

The instrument need to be at the ”Main Menu“ or press
Barcode Programs in the ”Main Menu“.

3.

Insert sample cuvette or blank (depending on the working
procedure) into the cell compartment.

4.

Press Options, More... and then Edit.

5.

A data overview is displayed, including the specifications of the
test that is to be revised. Compare the displayed data with the
data in the working procedure, and edit the displayed data in
accordance with the working procedure data.